Distortion measurements of sound pressure level generated by a pistonphone
A supplementary interlaboratory comparison on pistonphone calibration was performed between 2018 and 2020 under the auspices of the Inter-American Metrology System. Seven national metrology institutes took part in this comparison. One pistonphone was circulated among the participants
for measurement of the generated sound pressure level, frequency, total distortion + noise and total harmonic distortion. This paper presents the results of distortion measurements and analyzes them. From the stability check, notable variances with respect to reported measurement uncertainty
were observed. From reported results, a large difference between the estimated uncertainties by the participants was noted. Convergence between results was found when measurements were performed using one-inch microphone while some divergence between results was found when measurements were
performed using half-inch microphone. In addition, these results (using one-inch and half-inch microphones), even when obtained by the same institute, sometimes were different. Reasons for these findings are discussed and suggestions for future comparisons are presented.
Document Type: Research Article
Publication date: 01 August 2021
The Noise-Con conference proceedings are sponsored by INCE/USA and the Inter-Noise proceedings by I-INCE. NOVEM (Noise and Vibration Emerging Methods) conference proceedings are included. All NoiseCon Proceedings one year or older are free to download. InterNoise proceedings from outside the USA older than 10 years are free to download. Others are free to INCE/USA members and member societies of I-INCE.
- Membership Information
- INCE Subject Classification
- Ingenta Connect is not responsible for the content or availability of external websites
- Access Key
- Free content
- Partial Free content
- New content
- Open access content
- Partial Open access content
- Subscribed content
- Partial Subscribed content
- Free trial content